Analytical Industries PSR-11-917-M oxygen sensor for equipment that specifies this complete sensor model.
Key specifications
- Measurement range: 0–100% oxygen
- Signal output: 9–14 mV in air at 25°C and 1 atm
- Electrical connection: 3-pin Molex
- Response time to 90%: 13 seconds
- Operating temperature: 0–45°C
Model matching
The manufacturer's cross-reference lists PSR-11-917-M for Carefusion Avea and SiPAP, 300/900 ventilators with references 6419332 or 9004979, and Spacelabs reference 90514. Confirm the equipment service specification, full sensor suffix and connector before ordering.
PSR-11-917-M, M2 and MH1 are distinct models. The manufacturer lists Puritan Bennett 760/840 reference 4-072214-00 under PSR-11-917-M2, and Air Shields C200/C2000 reference 8362030 under PSR-11-917-MH1. Those are not confirmed applications for this M listing.
Send your sensor label and equipment model for a match check. Follow the equipment manufacturer's replacement and calibration procedure.
Technical references and expected life
The datasheet gives an expected life of 60 months in air at 25°C and 1 atm. Actual life depends on oxygen concentration, temperature and pressure; this is not a service-life guarantee.